Optical Power Meter (PM)

  • functionality check, optical detector cleaning, offset setting etc.
  • optical power calibration (CMC Uncertainty 2.8 %, 850 – 1625 nm wavelengths depends on use: e.g. 1310 / 1550 nm for standard PM for SM fibers)
  • linearity calibration (-4, -10, -20, -30, -40, -50, -60 dBm, CMC Uncertainty 0.14 dB)

Optical Light Source (LS)

  • functionality check, optical connector cleaning etc.
  • output power level in optical fiber (CMC Uncertainty 0.15 dB for Lasers)
  • spectral parameters (600 – 1650 nm): peak power wavelength (CMC Uncertainty 0.11 nm), FWHM
  • power stability (30, 120 min.)

Optical Loss Tester (LT)

  • functionality check, optical detector and connector cleaning, offset setting etc.
  • optical power calibration (CMC Uncertainty 2.8 %, 850 – 1625 nm wavelengths depends on use: e.g. 1310 / 1550 nm for standard PM for SM fibers)
  • linearity calibration (-4, -10, -20, -30, -40, -50, -60 dBm, CMC Uncertainty 0.14 dB)
  • output power level in optical fiber (CMC Uncertainty 0.15 dB for Lasers)
  • spectral parameters (600 – 1650 nm): peak power wavelength (CMC Uncertainty 0.11 nm), FWHM
  • power stability (30, 120 min.)
  • loss measurement calibration in automatic mode (with internal detectors)

Optical reflectometer OTDR

  • functionality check, optical connector cleaning etc.
  • peak power wavelength (CMC Uncertainty 0.11 nm)
  • optical distance / fiber length calibration
  • optical loss / linearity calibration (CMC Uncertainty 0.021 dB/dB)

Optical Spectral Analyser (OSA)

  • functionality check, optical connector cleaning etc.
  • wavelength scale calibration (depends on use and type: e.g. in DWDM band (1530 – 1560 nm); CMC Uncertainty 0.012 nm)
  • optical power scale calibration (CMC Uncertainty  0.16 dB)
  • linearity calibration (-5, -6, -7, -8, -9, -10, -15, -20, -30, -40, -50, -60 dBm, CMC Uncertainty 0.19 dB)

TV analysers

  • functionality check, mechanical check of the output connector and + cleaning
  • RF power calibration (-70 to +7.4 dBm, 5 – 2050 MHz)
  • linearity calibration – RF power / frequency calibration
  • instruments with auto-correction – setting calibration constants

We calibrate also other kinds of fiber optics instruments as e.g. Back-reflection Meter / Optical Return Loss Tester (ORL), Optical Attenuator (OA / VA) or Polarization Mode Dispersion Analyser (PMD) …